FCS Assay Development

Useful information to guide you through the steps of designing and implementing FCS experiments in solution.  This content was originally displayed on fcsxpert.com.

FCS Assay Development Steps

    Choosing an FCS Assay Type

      An overview of the interactions that can be measured with FCS and the FCS assay types used to monitor them.

        Selecting Control Samples for FCS Assays

        Helpful guidelines to aid you in achieving the most robust and meaningful results from your FCS assays using control samples.

        Preparing FCS Sample

          A guide to important sample conditions to consider when designing an FCS assay.

          Choosing FCS Acquisition Options

            A list of important measurement options to optimize when acquiring FCS data for your assay.

              Analyzing FCS Data

              See our Analysis Notes Section for guides and tips on analysis of experimental FCS data.

              Choosing an FCS Assay Type

              Fluorescence Correlation Spectroscopy (FCS) allows you to measure the diffusion, concentration and molecular interactions of any fluorescent molecules or particles.

              These guides will help you understand the general types of solution-based FCS assays, which can be performed on our QuantumXpert FCS spectrometer, and the types of FCS assays that are best suited to measure the interactions that interest you.

              FCS Assay Types Guides

              • Biological Interactions that FCS Can Detect

              Descriptions of the different categories of molecular interactions that can be measured with FCS.

              • Autocorrelation vs. Cross-correlation Assays

              The properties of the two correlation assay types when measuring different forms of molecular interactions and critical factors to consider when designing and implementing these FCS measurements.

              Selecting Control Samples for FCS Assays

              In order to successfully analyze FCS data, several control measurements should be made to verify you results. The precautions outlined below will ensure that your FCS analysis yields unambiguous results.

              Measure the buffer alone

              This will give you information on the background signal from the buffer, or any buffer additive (e.g. BSA or detergent) present in the sample.

              Measure each dye alone in buffer

              This will give you useful information on the diffusion time of free dye, and will help you interpret autocorrelation data should there be free dye contaminating your labeled probe.

              Measure each dye in the presence of each unlabeled probe

              This will determine if a dye non-specifically interacts with one of the probes being used.

              Measure each labeled probe alone in buffer

              This will give you useful information on the diffusion time of labeled probe and will help you decide the effective concentration range that can be used to give optimal fluorescence intensity (counts per second).

              Preparing FCS Samples

              Fluorescence Correlation Spectroscopy offers exceptional fluorescence detection sensitivity. High sensitivity measurements require carefully optimized measurement conditions to ensure that you are not characterizing artifacts of the sample.

              Sample Optimization Guides

              The AssayXpert Kits currently being developed by Sensor Technologies LLC will provide you with quality controlled reagents to ensure that your FCS measurements with the QuantumXpert are free of artifacts.

              Choosing FCS Acquisition Options

              FCS data acquisition is very simple, but choosing the wrong options can lead to ambiguous or complicated results. Carefully considering the following options will ensure that you acquire meaningful data.

              Laser excitation intensity

              Must be set to a suitable intensity for your fluorescent dyes to avoid photobleaching and intersystem crossing (triplet state excitation), yet achieve sufficient average intensity for FCS measurements.

              Delay Before Acquistion

              The time between sample preparation and measurement depends on the type of assay being performed. Assays that monitor reaction dynamics over time should be measured as soon as possible, but assays that monitor systems at equilibrium should be measured when you are sure equilibrium has been reached. A significant delay between preparation and measurement can lead to aggregation of your sample and can affect some fluorescent dyes.

              Acquisition Duration

              Longer data acquisition leads to cleaner, more robust data. However, very long acquisition times increase the chance that your data will be affected by artifacts such as large fluorescent particulates or drift in intensity over time.